Additel 223A

Category:

Description

OVERVIEW

A highly integrated multifunction calibrator featuring several patented technologies, the 223A is an ultra-compact, rugged, and easy to  use hand-held device for sourcing, simulating and measuring pressure, temperature, and electrical signals. Its smartphone-like menu and interface make the operation simpler and easier. With HART communication capability, the 223A is ideal for calibrating, maintaining, and troubleshooting HART instrumentations. Automation and documentation capabilities make the 223A a turnkey solution.

 

FEATURES

Sourcing, simulating and measuring pressure, temperature and electrical signals

• Smartphone-like menu and interface make the operation simpler

• Hart Communication capability

• The internal cold junction compensation sensor can be re-calibrated at ice point by user

• Ultra-compact, 3.9” x 7.6” x 2.0”, and 1.6 lb (0.7 kg)

 

APPLICATIOS

The 223A multifunction calibrator is a process tool for measuring, sourcing and simulating mA, mV, V, RTDs, thermocouples, ohms, frequency, pulses and pressure, captures switch values and provides 24 V loop power.
• Electrical
  Resistance measurement / simulation
  Voltage measurement / generation
  Current measurement /generation
  Frequency measurement / generation
  Pulse counting/generation
  24 VDC loop supply
  Switch sensing
• Temperature
  RTD measurement /simulation
  TC measurement /simulation

• NIST Traceable Calibration with data

• Warranty:
    3 years
SPECIFICATIONS

Measurement Accuracy
Range Resolution Accuracy
Voltage DC ±75.0000 mV 0.1μV 0.01%RD + 3.75 μV
±30.0000 V 0.1 mV 0.01%RD + 1.5 mV
Current DC ± 30.0000 mA 0.1μA 0.01%RD + 1.5 μA
Resistance Two-wire 0 to 400.000 Ω 1mΩ 0.02%RD + 0.02 Ω
Three-wire 0 to 400.000 Ω 1mΩ 0.02%RD + 0.02 Ω
Four-wire 0 to 400.000 Ω 1mΩ 0.01%RD + 0.02 Ω
Two-wire 0 to 4000.00 Ω 10mΩ 0.02%RD + 0.2 Ω
Three-wire 0 to 4000.00 Ω 10mΩ 0.02%RD + 0.2 Ω
Four-wire 0 to 4000.00 Ω 10mΩ 0.01%RD + 0.2 Ω
Frequency 1 to 50000.0 Hz 0.1Hz 0.005%RD + 1 Hz
Pulse 0 to 999999 1 N/A
Limit Switch For the contact with potential, the voltage within the range 3V to 24V.

 

Source Accuracy
Voltage DC -10.000 to 75.000mV 1μV 0.02%RD + 4.25 μV
0 to 12.0000 V 0.1mV 0.02%RD + 0.6 mV
Current DC 0 to 22.000 mA 1μA 0.02%RD + 1.1 μA
Resistance 1 to 400.00 Ω 10mΩ 0.02%RD + 0.02 Ω
1 to 4000.0 Ω 100mΩ 0.03%RD + 0.4 Ω
Frequency 0 to 50000.0 Hz 0.1Hz 0.005%RD + 1 Hz
Pulse 0 to 999999 1 N/A
DC24V N/A N/A 0.5V

 

Thermocouple Measurement and Source Accuracy
Measure and Simulate Standard Temperature Range (oC) Accuracy (oC)
Measure Source
S IEC 584 -50 to 1768 -50 to400 1 1.1
400 to 1000 0.6 0.6
1000 to1768 0.7 0.8
R IEC 584 -50 to 1768 -50 to 200 1.4 1.4
200 to 500 0.6 0.6
500 to 1768 0.6 0.7
B IEC 584 0 to 1820 50 to 450 3.8 3.8
450 to 800 0.9 0.9
800 to 1820 0.6 0.7
K IEC 584 -270 to 1372 -250 to -200 1 1.1
-200 to -100 0.4 0.5
-100 to 600 0.3 0.3
600 to 1372 0.4 0.5
N IEC 584 -270 to 1300 -250 to -200 1.5 1.6
-200 to -100 0.5 0.6
-100 to 1300 0.4 0.5
E IEC 584 -270 to 1000 -250 to -200 0.6 0.7
-200 to -100 0.3 0.3
-100 to 0 0.2 0.2
0 to 700 0.2 0.3
700 to 1000 0.2 0.4
J IEC 584 -270 to 1200 -210 to -100 0.3 0.3
-100 to 1200 0.3 0.4
T IEC 584 -270 to 400 -250 to -200 0.8 0.9
-200 to 0 0.4 0.4
0 to 400 0.2 0.2
C ASTM E988 0 to 2315 0 to 1000 0.5 0.5
1000 to 1800 0.7 0.9
1800 to 2315 1 1.4
D ASTM E988 0 to 2320 0 to 100 0.5 0.5
100 to 1100 0.4 0.5
1100 to 2000 0.6 0.9
2000 to 2320 0.9 1.3
G ASTM E1751 0 to 2315 0 to 200 2.4 2.4
200 to 400 0.5 0.5
400 to 1400 0.4 0.5
1400 to 2315 0.7 1
L DIN43710 -200 to 900 -200 to -100 0.2 0.3
-100 to 400 0.2 0.2
400 to 900 0.2 0.3
U DIN43710 -200 to 600 -200 to 0 0.4 0.4
0 to 600 0.2 0.3
*Accuracy with external cold junction; for internal cold junction add 0.1°C (k=2)

 

Measurement Accuracy
Measure and Simulate Standard Temperature Range (°C) Accuracy (°C)
Measure (2W/3W) Measure (4W) Source
Pt10(385) IEC 751 -200 to 850 -100 to 200 0.65 0.6 0.65
200 to 600 0.82 0.72 0.82
600 to 850 0.96 0.82 0.96
Pt100(385) IEC 751 -200 to 850 -100 to 200 0.15 0.1 0.15
200 to 600 0.26 0.16 0.26
600 to 850 0.34 0.2 0.34
Pt100(3916) JIS 1604 -200 to 850 -100 to 200 0.15 0.1 0.15
200 to 600 0.26 0.16 0.26
600 to 850 0.33 0.2 0.33
Pt100(3926) Minco Application Aid #18 -200 to 850 -100 to 200 0.15 0.1 0.15
200 to 600 0.26 0.16 0.26
600 to 850 0.33 0.2 0.33
Pt500(385) IEC 751 -200 to 850 -100 to 200 0.2 0.16 0.36
200 to 600 0.32 0.22 0.54
600 to 850 0.4 0.27 0.67
Pt1000(385) IEC 751 -200 to 850 -100 to 200 0.1 0.05 0.25
200 to 600 0.2 0.1 0.42
600 to 850 0.27 0.14 0.54
Cu10(427) Minco Application Aid #18 -100 to 260 -100 to 260 0.61 0.56 0.61
Cu50(428) GOST 6651-94 -50 to 150 -50 to 150 0.17 0.13 0.17
Cu100(428) GOST 6651-94 -50 to 150 -50 to 150 0.12 0.09 0.12
Ni120(672) Edison curve #7 -100 to 260 -100 to 260 0.07 0.05 0.07
Ni100(618) DIN 43760 -100 to 260 -100 to 260 0.08 0.06 0.08

 

General Specifications
Operating Temperature -10°C to 50°C
Storage Temperature -20°C to 60°C
Humidity <90%, non-condensing
European Compliance CE Mark
Display 3.5 inch TFT color screen
Electrical Connection Ø4mm sockets and flat mini-jack thermocouple socket
RS232 Interface Standard RS232-DB9 socket
Size 3.9″ x 7.6″ x 2.0″ (100mm x 192mm x 52mm)
Weight 1.6 lb (0.7 kg)
Power Supply Polymer Li-ion rechargeable battery,or 10V DC adaptor
Battery Rechargeable Li-ion battery (included)
Battery Life 15 hours uninterrupted use
Battery life will be reduced when 24V is applied
Battery Charge 110V/220V external power adapter (included)
Pressure Specifications (222A & 223A) The 160A series Intelligent Digital Pressure Modules are available for gauge, vacuum and absolute pressure from -15 psi to 10,000 psi (-1 bar to 700 bar). Accuracy from 0.02% FS includes operation over 14°F to 122°F (-10°C to 50°C), one year stability and calibration uncertainty. For detail specification refer to pressure modules datasheet.

 

Reviews

There are no reviews yet.

Be the first to review “Additel 223A”

Your email address will not be published. Required fields are marked *